Serving Scientists Since 1994

USP5 Watch Glass Transdermal Patch Holder with 17 Mesh Screen (Paddle Over Disc)

Part Number Pack Size Price Qty
APPFIVE-HR Each

USP5 Watch Glass Transdermal Patch Holder with 17 Mesh Screen


OEM Part# 65-190-093

Apparatus

This QLA USP 5 apparatus for transdermal testing is sold as a borosilicate watch glass with a PTFE 17 mesh, held together by PTFE clips. Patches up to 90mm in diameter can be tested. It is compatible with all brands of dissolution bath and conforms exactly to the USP specification.

 

Method

Transdermal or patch testing is carried out using USP method 5 (paddle over disc) or USP method 6, the rotating cylinder.

With paddle over disc, the transdermal patch is placed between a glass disc and an inert PTFE mesh. This is placed at the bottom of the vessel, with the mesh facing upwards, under a rotating paddle.

Unlike dissolution testing, transdermal testing is carried out at 32°C to reflect the lower temperature of the skin. Other variables such as the height setting and sampling requirements are the same as dissolution testing.

Hanson Part 65-190-093

Pack Size Each
Transdermal USP5 Disc

Product Tags