The Q-Grad Pump provides high-pressure quaternary gradients for HPLC applications. Four high-performance pumps are packaged into a single cabinet and are controlled with PC-based software. The result is high-pressure gradient capability that is as economical as low pressure mixing systems.
The Q-Grad Pump uses a very efficient pulse damper mixer to ensure high performance gradients. A prime-purge valve and upper and lower pressure limits are also standard. All Q-Grad pumps carry a 3-year warranty.
External synchronization from an injector or autosampler may be enabled through the pump control option using standard TTL or contact closures.
1. Stand-alone keypad (System Control Unit SCU-460)
2. EZStart system control and data acquisition software controls of the Q-Grad pump, detectors, and autosampler.
Both have Run/Stop inputs and real-time pressure monitoring with upper and lower pressure limits. Each option is designed to be user friendly. You choose the interface that fits your needs.
High pressure mixing of solvents eliminates the common problem of solvent outgassing, pre-pump mixing volume, and differing solvent compressibilities associated with low pressure mixing.
• Self Flushing Pump Heads
• Prime-Purge Valve
• Pulse Damper for Reduced Pulsation
• Stainless Steel or PEEK Fluid Path
• Motor Stall Detector
• Spring Loaded Check Valves
Standard Pump Controls:
• RS-232 Serial COM Port (each pump)
• Run/stop Input (each pump) - 5 volt TTL type
• Real Time Pressure Reading
• 0-100% in 0.1% increments
• Percents B, C, D programmable
• Percent A automatically adjusted
| Specifications | |
| Flow Rates: | 0.1 - 5.0ml/min (each pump head) |
| Flow Accuracy: | ±2% |
| Flow Precision: | 0.2% RSD |
| Pressure: | 0 - 6000psi (stainless steel) 0 - 5000psi (PEEK) |
| Pressure Accuracy: | ±1% of full scale pressure |
| Pulsation: | ±1% at 1ml/min and 1000psi |
| Physical Characteristics | |
| Dimensions (W x H x D): | 264 x 190 x 445mm |
| Mass: | 16kg |
| Power: | 110-220V AC, 50/60Hz |